5.1 NORMAL OPERATION – Temperature and relative humidity are continuously displayed.
The asterisk “*” adjacent to the temperature reading shows when the heater power is on. When warming
the asterisk will be continuously on, once warmed up the asterisk will slowly flash as the heater is pulsed
to maintain the correct temperature. When reducing the temperature setting the asterisk may go off, this is
normal.
The asterisk “*” adjacent to the relative humidity display is only on when the pump control output is on
(see section 8) and is only applicable when using the optional Brinsea Advance Humidity Management
Module.
5.2 POWER LOSS DISPLAY – If power has been interrupted due to a power cut (or when first switching
on) a “P” is shown flashing in the corner of the display. Press OK for 2 or more seconds to clear the
indicator. If the reason for the power loss is not known check the power cable connections are secure.
Once the “P” indicator has been cleared, it is advisable to candle eggs a number of times to check for
losses.
5.3 HIGH TEMPERATURE ALARM DISPLAY – If the measured temperature goes up by more than the
figure in the ALARM HI screen, the alarm will sound immediately and “+T” will be displayed. Press OK
to silence the alarm for 30 minutes.
If the high temperature problem rectifies itself the “+T” remains on the display to show this has
happened. Press OK to clear the indicator. Check the incubator is not (and has not been) in direct sunlight
or too near a heat source such as a room heater. It is advisable to candle eggs a number of times after this
event to check for losses.
5.4 LOW TEMPERATURE ALARM DISPLAY – If the measured temperature goes down by more than the
figure in the ALARM LO screen, after 30 minutes “-T” will be displayed and the alarm will sound. Press
OK to silence the alarm for 30 minutes.
If the low temperature problem rectifies itself the “-T” remains on the display to show this has happened.
Press OK to clear the indicator. Check the incubator is not (and has not been) in a cold draught or that the
room temperature has dropped significantly. It is advisable to candle eggs a number of times after this
event to check for losses.
T*37.5C
H 45% P
T 39.8C
H 45% +T
T*32.1C
H 45% -T